® DirectLine DL424/425 Sensor Module for DL 5000 Dissolved Oxygen Probes User Manual 70-82-25-113 Rev. 3 September 2011 Honeywell Process Solutions...
Page 2
September, 2011 Warranty/Remedy Honeywell warrants goods of its manufacture as being free of defective materials and faulty workmanship. Contact your local sales office for warranty information. If warranted goods are returned to Honeywell during the period of coverage, Honeywell will repair or replace without charge those items it finds defective.
Page 3
Sensor. Insert 70-82-10-04, a quick reference guide for configuring and calibrating the DL424/425, should accompany this document. Contacts World Wide Web The following lists Honeywell’s World Wide Web sites that will be of interest to our customers. Honeywell Organization WWW Address (URL) Corporate http://www.honeywell.com...
Introduction 1. Introduction 1.1 Overview The DirectLine® Sensor for dissolved oxygen measurement in water consists of a DL424/425 electronics module connected to a DL5000 DO Probe. The modular electronics design allows the module to be separated from the probe, so that the probe can be easily removed or replaced while retaining power to the electronics module.
Introduction 1.2 Electronics Module The electronics module is loop-powered by 16-42 Vdc and will modulate its supply current from 4 mA to 20 mA, depending upon the DO value that is measured by the probe. The transmitted loop current is compensated for process temperature. A 4-20 mA output connection is provided via a 6m cordset or a customer supplied cable used in combination with a field wiring connector.
Introduction 1.4 Specifications General: Displayed Process DL424: 0-20 ppm dissolved oxygen Variable DL425: 0-200 ppb dissolved oxygen Displayed 2.0 to 60.0 ºC (35.6 to 140ºF) Temperature Range Operating –20 °C to +60 °C (-4 °F to +140 °F) Temperature Storage Temperature –20 °C to +70 °C (-4 °F to +158 °F) Display Resolution DL424: 0.01 ppm...
Introduction 1.5 Model Selection Guide Instructions Select the desired key number. The arrow to the right marks the selection available. Make the desired selections from Tables I through IV using the column below the proper arrow. A dot ( ) denotes availability. Key Number _ _ _ _ _ _ _ _ _...
Installation 2. Installation 2.1 Assembly and Wiring Depending on the customer selected output cable options, the DirectLine can be wired to an appropriate 16-42 Vdc source using 2 different methods: 1) Cordset. See Figure 2-1 Cordset Connection and Wiring 2) Field wiring connector with customer supplied cable. See page 6. Refer to Section 7 for wiring for CE Mark applications.
Installation 2.1.2 Field Wiring Connector with customer supplied cable Refer to Figure 2-2. The field wiring connector supports customer supplied cable with an outer diameter of 4-6mm, 2-wire twisted shielded pair. Table 2-1 Assembly and Wiring Procedure for Field Wiring Connector Step Procedure Disassemble field wiring connector...
Installation 2.2 Integral Mounting Table 2-2 Integral Mounting Procedure (refer to Figure 2-3) Step Procedure Connect the probe to the process source (using the appropriate mounting from those supplied for the DL5000). Make sure that the final position of the installed electronics module allows the display to be easily viewed by personnel.
Installation 2.3 Remote Mounting When the DL424 or DL425 module is specified with Table II = 6, a remote connector assembly (part number 51500768-005) is supplied loose. The remote cable connector is used to connect the DL5000 probe cable to the DL424/425 module. Table 2-3 gives the mounting procedure.
Installation Hose Clamp Slot for Pipe Mounting (Hose Clamp included in kit) Through Hole for DIN Rail Mounting Metal clip for Gray clip for “U” DIN rail mounting “G” DIN rail mounting Mounting Kit View from End View from End Panel Panel Panel...
Installation 2.4 Conduit connections The DirectLine provides a male ½” NPT thread to accommodate a customer conduit connection. Use ½” conduit coupling (min. 38.1mm (1.5”) long) on DL conduit connection to clear cordset connector. Conduit can not be used with field wiring connector (see Figure 2-2) due to size restriction.
Configuration 3. Configuration 3.1 Overview Configuration Parameters Set Up consists of configuring the following functions: Atmospheric Pressure Compensation: Table 3-1 Dependence of Barometric Pressure on Altitude The solubility of oxygen in water and DL5000 dissolved oxygen probe signal directly respond to barometric pressure and hence pressure compensation is Alt, ft P, mmHg...
Page 20
Configuration Probe Bias Voltage – The probe bias voltage value determines the voltage applied to the DO probe's cathode, relative to the reference electrode, to drive the reduction of oxygen to water at the cathode and thereby induce a current which is directly proportional to the oxygen consumed.
Configuration 3.2 Configuration Set Up Procedure ATTENTION: In Table 3-2, under the Press column: Hold means to hold the button down until the display changes. Momentarily means to press and release the indicated button. From the Online DO display, follow this procedure. ATTENTION: If no key is pressed for 60 seconds, the display will abort the entry mode and default to Online Display.
Page 22
Configuration Step Operation Press Display Enter Probe Bias MODE b IAS (for 1 second), then Voltage Momentarily Value of current Probe Bias Voltage in Volts Go to 3a, 3b, or 3c Remove the probe from the process prior to 3a - Probe Bias Calibration. Flashing Display –...
Page 23
Configuration Step Operation Press Display Enter Output MODE OutC Enter Output Calibration Configuration Momentarily 100% Range Value rnGH (for 1 second) then, Selection (Value of current 100 % Range Value Momentarily Selection) Edit 100% Range MODE Flashing Display – You are now in EDIT Value Selection Hold mode...
Page 24
Configuration Step Operation Press Display 0 % Calibration AdJL Momentarily Adjust 0 % MODE AdJL (flashes) – You are now in EDIT Calibration Hold mode Range: 3.80 to 4.40 mA typically (default 4.00 mA) +AdJL (increments value) AdJL (decrements value) Momentarily Save 0 % MODE...
Calibration 4. Calibration 4.1 New Probe New Probe Reset Option Upon system power up or probe connection, the new probe reset option is displayed. The nPrb message is available for 2 minutes and can be selected by pressing the up arrow until the display changes to the unit type and DO value. The Air and Sample calibration values are set to the factory defaults and the probe current is measured and saved as a diagnostic aid.
Calibration 4.3 Calibration Procedures ATTENTION: In Table 4-1, under the Press column: Hold means to hold the button down until the display changes. Momentarily means to press and release the indicated button. Table 4-1 Calibration Procedure Step Operation Press Display ACAL...
Page 27
Calibration Step Operation Press Display Sample Calibration SCAL (1 second), then "Live DO Value" SCal Hold To cancel, press Mode button (3 seconds) The display changes to a live DO reading, SCal so you can continue to monitor the sample. Momentarily...
Operation 5. Operation 5.1 Displays Overview The DirectLine® DL424/425 displays the On-line DO Concentration value and the On- line Temperature. The table below describes these parameters. Table 5-1 Online Parameter Descriptions Parameter Description Online DO D424 Range: ppm = 0.00 to 20.00 D425 Range: ppb = 0 to 200 Online Measured temperature expressed with fixed tenths decimal...
Operation ATTENTION: In Table 5-2, under the Press column: Momentarily means to press and release the indicated button. Table 5-2 Display Navigation Procedure Step Operation Press Display View Online DO MODE (measured DO) Concentration value Momentarily View Online MODE (measured temperature in °C or °F) Temperature Momentarily...
Operation 5.2 Diagnostic Error Messages When a diagnostic error or status condition occurs, the Online Display alternates between measured DO and a text message. Table 5-3 Online Diagnostic Errors What you see Cause of Error What to do CnFG Configuration or Calibration data is Reset unit or cycle power.
Page 31
Operation What you see Cause of Error What to do FAIL This error can occur if Verify that Probe has been removed from the process. The probe current is too low or if the probe readings are unstable during air calibration.
Operation 5.3 Unit Reset Overview ® Unit Reset initializes all of the DirectLine Sensor’s calibration and configuration data to factory default values. The Air calibration value is also reset to the factory default. Procedure From the Online DO display, press and hold the buttons simultaneously until “rSEt”...
Field Wiring connector The Field Wiring Connector is an all-plastic screw terminal M12 female type that can be purchased directly from Honeywell or from multiple vendors including: Turck Industries Part Number B8141-0 for a M12 field wiring connector that accommodates customer supplied cable.
Page 34
Spare Parts DirectLine® DL424/425 Sensor Module User Manual 9/11...
7. Appendix: CE Mark Applications CE Conformity (Europe): This product is in conformity with the protection requirements of 89/336/EEC, the EMC Directive. Conformity of this product with any other “CE Mark” Directive(s) shall not be assumed. Deviation from the installation conditions specified in this manual, and the following special conditions, may invalidate this product’s conformity with the EMC Directive.
Honeywell A/S Brazil SE Asia Countries FAX: +971 2 4432536 Fax: +(86-29) 8833-7489 Phone: (45) 39 55 55 55 Honeywell do Brasil & Cia see Honeywell Sultanate of Oman Phone: +(55-11) 7266-1900 Automation India Ltd for: China – PRC - Shenzhen- Poland Honeywell &...
Page 37
Appendix: CE Mark Applications 9/11 DirectLine® DL424/425 Sensor Module User Manual...
Page 38
Honeywell Process Solutions 1100 Virginia Drive 70-82-25-113 Rev.3 Fort Washington, PA 19034 September 2011 www.honeywell.com/ps 2011 Honeywell International Inc.