Important Information — Please Read Before Use Intended use The Olympus WM-DP1 mobile workstation is intended for use in medical facilities under the direction of a trained physician, and has been designed to be used with a range of Olympus equipment to facilitate GI endoscopy, endoscopic ultrasound, respiratory and surgical endoscopic procedures.
Standard Set and Accessories Chapter 1 Standard Set and Accessories Please refer to the standard set below and check that all items are present. Contact Olympus if any parts are damaged or missing. Workstations that are fitted with antistatic castors can be identified by the following label affixed to the rear of each shelf.
Shelf kit narrow MAJ-198 Shelf kit wide MAJ-199 Universal display arm CRT MAJ-1613 – – Universal stowage container MAJ-185 Videoscope cable holder MAJ-1630 compatible – incompatible CAUTION Do not attempt to install any other accessories on the WM-DP1. MOBILE WORKSTATION WM-DP1...
Warnings and Cautions Chapter 2 Warnings and Cautions WARNING • The WM-DP1 mobile workstation should only be used in a medical facility under the direction of a trained physician. • Explosion hazard - never install or use the WM-DP1 mobile workstation within the zone of risk of flammable gases.
Page 7
• Section 6.2 Manoeuvring and Transportation gives detailed instructions and precautions which must be read and understood before attempting to move the mobile workstation. • Spray type medical agents, such as lubricants, anaesthetic or alcohol, should be used away from the mobile workstation. MOBILE WORKSTATION WM-DP1...
Page 8
• Store and use the WM-DP1 mobile workstation within the environmental conditions described in Chapter 8; failure to do so may lead to equipment malfunction or failure.
IEC 60085:2004 Unit complies with relevant Disposal conditions - see End European directives of Life (Chapter 12) (MAJ-175 only) (MAJ-175 only) View of the separation transformer rear panel NOTE MAJ-175 (220-240V) is shown. Other versions are similar. MOBILE WORKSTATION WM-DP1...
• Multiple fixing positions are provided enabling the shelves to be raised or lowered to accommodate varying equipment configurations (see 4.2). • The mobile workstation is designed for use with Olympus OEV LCD monitors and other monitors that meet the VESA standard. Contact Olympus regarding compatibility of other monitors.
Page 11
Additional fixings are provided (Figure 4.1) to facilitate shelf adjustment or the fitting of the optional drawer pack(s). MOBILE WORKSTATION WM-DP1...
Page 12
Preparation for Use 318.00 Figure 4.1 MOBILE WORKSTATION WM-DP1...
Preparation for Use 4.2 Shelf positions NOTE The WM-DP1 mobile workstation is supplied with the shelves fitted as shown in Figure 4.1. It is recommended that shelves are repositioned, if required, before commencing equipment installation as described below. Position the mobile workstation on a flat surface and apply both castor brakes (diagonally opposed).
The mobile workstation is supplied with the remote switch connected to the separation transformer and no assembly is required. However, check the connection (see Figure 4.3) is secure before completing full installation. Potential equalisation terminals Fuse Remote switch connector Figure 4.3 MOBILE WORKSTATION WM-DP1...
Place items of equipment centrally on the shelving and connect according to individual instructions. The WM-DP1 mobile workstation is supplied with a universal IEC lead set to connect equipment on the mobile workstation to the transformer. Potential equalisation terminals are fitted to both sides of the transformer enclosure for connection to equipment potential equalisation terminals where fitted.
Page 16
LED will illuminate yellow indicating ‘standby’. Set the mobile workstation remote switch to the ON position, the LED will now illuminate green indicating ‘output on’. Set the power switches for the installed equipment to the ON position. MOBILE WORKSTATION WM-DP1...
Blanking cap, monitor port Blanking cap, cable port Handwheel Figure 4.5 The cable hanger locates into either left or right hand port in the top tray and is secured in the required position using handwheel as shown in Figure 4.5. MOBILE WORKSTATION WM-DP1...
Preparation for Use 4.6 Water container holder The WM-DP1 mobile workstation is fitted with a water container bracket to hold two water containers (not supplied). The water container simply clips onto the bracket as shown in Figure 4.6. Figure 4.6 4.7 Rear panels...
Preparation for Use 4.8 Fitting the MAJ-181 restrictor This procedure is only applicable if fitting a MAJ-181 LCD monitor arm to the WM-DP1 workstation. WARNING • If using a MAJ-181 on the workstation, the restrictor supplied in this kit must be fitted to the MAJ-181 and is best assembled with the MAJ-181 in position on the workstation.
Page 20
M8 x 50mm screw. Fully tighten the screw using the M6 A/F Allen key supplied. NOTE The monitor arm must be parallel to the top tray to ensure the tag on the monitor arm aligns with the cutout in the restrictor (Figure 4.10 inset). MOBILE WORKSTATION WM-DP1...
Page 21
Preparation for Use MAJ-181 restrictor - rear M8 x 50mm screw MAJ-181 restrictor - front M8 coil washer M8 plain washer Figure 4.10 Repeat for second MAJ-181 if used. MOBILE WORKSTATION WM-DP1...
Where antistatic castors are fitted, periodically check the electrical resistance of the conductive antistatic castors to assess antistatic effectiveness. Check the security of castors at six-monthly intervals. If loose, contact Olympus for service. Regularly check the shelf fixing screws and the fixing screws of any accessories that are fitted, particularly monitor mounts.
Chapter 6 Operation 6.1 Operation WARNING • The WM-DP1 should be used in accordance with the following. Otherwise, the workstation will be ‘top heavy’ and may lose stability and topple, causing equipment damage and/or personal injury. - Two MAJ-181 restrictors are supplied which must be fitted to MAJ-181 LCD monitor arm(s) if used (refer to section 4.8).
Page 24
Operation Figure 6.1 After use, turn off the remote switch. MOBILE WORKSTATION WM-DP1...
Operation 6.2 Manoeuvring and Transportation WARNING • The WM-DP1 should be manoeuvred or transported in accordance with the following. Otherwise, the workstation may lose stability and topple, causing equipment damage and/or personal injury. - Two people are required to safely manoeuvre or transport a loaded mobile workstation.
Page 26
Operation Higher level Higher level Lower level Lower level Figure 6.2 NOTE Repeated transportation over uneven surfaces may result in damage to the equipment carried on the mobile workstation. MOBILE WORKSTATION WM-DP1...
During cleaning, wear appropriate protective equipment, such as eye wear, face mask, moisture-resistant clothing and chemical- resistant gloves that fit properly and are long enough so that your skin is not exposed. MOBILE WORKSTATION WM-DP1...
The WM-DP1 mobile workstation should be stored within the environmental conditions given in Chapter 8. 7.3 Maintenance Check the security of castors at six-monthly intervals. If loose, contact Olympus for service. Where antistatic castors are fitted, periodically check the electrical resistance of the conductive antistatic castors to assess antistatic effectiveness.
NOTE: Equipment stored on the mobile workstation should not exceed the size of the shelf. Castors WM-DP1 p/no K10003727 4 x 125mm antistatic, two with brakes. WM-DP1 p/no K10003726, K10003728, K10003729 4 x 125mm non-antistatic, two with brakes. Ambient temperature Storage: -40°C to +70°C (-40°F to +158°F) Operation: 10 - 40°C (50 - 104°F)
Page 30
Refer to your local Olympus distributor for return and/or collection systems available in your country. The Olympus WM-DP1 mobile workstation and separation transformer is manufactured in the UK by KeyMed (Medical & Industrial Equipment) Ltd. MOBILE WORKSTATION WM-DP1...
This can best be guaranteed by using only those spare parts and components specified and supplied by Olympus. If any item becomes damaged or lost, contact Olympus for replacement. MOBILE WORKSTATION WM-DP1...
Potential equalisation terminals are provided adjacent to the input cable and either side of the top cover, for use where local regulations require the enclosures of each part of the system be connected to a common earth point, independently of the mains supply cables. MOBILE WORKSTATION WM-DP1...
Page 33
Technical Description Figure 10.1 MOBILE WORKSTATION WM-DP1...
As above, plus investigate from other systems. source of interference. If the problem cannot be resolved, contact Olympus for service. There are no user-serviceable parts inside the separation transformer. This equipment must not be modified without written authorisation from Olympus.
End of Life Chapter 12 End of Life Equipment which has reached the end of its useful life should be disposed of according to local regulations, refer to section 8.2 for additional information. MOBILE WORKSTATION WM-DP1...
Page 36
491B, River Valley Road #12-01/04, Valley Point Office Tower, Singapore 248373 Fax: 6834-2438 Telephone: 6834-0010 OLYMPUS (BEIJING) SALES & SERVICE CO,. LTD. Room 1202, NCI Tower, A12 Jianguomenwai Avenue, Chaoyang District, Beijing, 100022 P.R.C. Fax: (10) 6569-3545 Telephone: (10) 6569-3535 OLYMPUS MOSCOW LIMITED LIABILITY COMPANY 117071, Moscow, Malaya Kaluzhskaya 19, bld.