(KS or ES), enter parameters used in the calculation (such as mud density, oil density), log results, view results and download previous records. Records may be exported as spreadsheets. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
3 seconds until the Power Off message ap- pears then release. Please ensure the unit is turned off after use to ensure the battery is not drained during storage. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Base Density: specific gravity of the base fluid • Oil Calibrated: for informational purposes only since all devices are always calibrated with diesel by default or user-selected base fluids OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 6
In the second stage, the software will prompt to run this stage using the diluted mud sample. This requires mixing 20 mL of whole drilling fluid with 205 mL of the DSX7 surfactant solution (#365-20-4). OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 7
PC and is now available for selection in the “New Measurement” window. The win- dow will automatically close once “OK” is clicked. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Power / Scroll / Hold to Cancel or Power Off Navigate Down One Level in Menu / Enter Increase or Decrease Parameter Value / Hold Both to Go Back One Level OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Ensure these are correct and use the enter key to continue. Mud Density can be adjusted on the device using the navigation and +/- keys. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 10
11. If operating the MudChecker v1.5, press the enter key to calculate and recall your measurement results. MudChecker v1.0 models require the device to be connected to the software to recall results. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
6. Once testing is completed, the software displays a confirmation message. The user then has the option to discard the test results or save the test results into the Test Archive. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Allen key through the two holes at the bottom end of the shield and applying force gently. Do not over tighten the probe shield. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Display test data and analysis and data interpretation • Report test results and basic drilling fluid properties • Archive test data for further analysis • Store device calibration data OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
PC, test data is automatically downloaded from the device and analyzed in the software. Analysis results and Excel-based reports can then be generated from the software. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 15
(for example FTDI) to download the required USB drivers for the PC. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
3. Click Next to select the installation location. Make sure the “Everyone” checkbox is checked. This ensures any users of the host PC will be able to access the software. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 17
The .NET Desktop Runtime software installation file is included on the thumb drive that is shipped with the MudChecker. If you prefer to use that file, skip to step 11. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 18
PC. The image above shows how the file appears in Microsoft Edge. In Google Chrome, the downloaded files are usually shown on the lower left-hand corner of screen as shown below. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 19
Double-click the MudCheckerApp icon on the desktop to run the software. 13. If the software is installed correctly, the main screen should display. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
The Action Bar allows the user to Run a New Measurement (Test), or to print an Excel-based report for any archived test. New Measurement Click the “New Measurement” button to initiate a new measurement (or test). OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Note that tests once downloaded from the device are removed from the device memory and archived in the host PC database. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 22
Click on any row in the data grid to view test inputs data and associated re- sults of the solids analysis calculations. Input data of the drilling fluid is shown in the “Test Input” group. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 23
LGS % / (lb/bbl): the volumetric fraction of low gravity solids (drill solids) in the drilling fluid sample, and the mass concentration in lb/bbl OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Any archived test available in the host PC can be exported as an Excel- based report. The various groups in the report are similar to those described Reports for the software. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
When carrying out a measurement ensure that all air bubbles are expelled. We recommend that you do this by giving the probe a gentle wiggle while submerged in the sample. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Damage to the unit enclosure or top lid may significantly degrade the IP rating of the unit. Such damage should be discussed with OFITE. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Page 28
If this does occur, please contact OFITE. • At lower temperatures, the internal battery will become discharged sooner. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
In the event that OFITE is requested to provide customized research and development for the buyer, OFITE shall use its best efforts but makes no guarantees to the buyer that any products will be provided. OFITE makes no other warranties or guarantees to the buyer, either express or implied, and the warranties...